TE ION CHAMBER
MEASURES RADS
GAMMAS
AND NEUTRONS
|
|
The
model 1030 is a Pulse Survey Meter. It has been optimized for use
in pulsed radiation fields and will measure absorbed dose from any
penetrating radiation including x-rays, gamma rays, neutron and
high energy particles. It has both rate and integrate ranges.
The detector is a cylindrical parallel plate ion chamber with interleaved
collection and signal plates. This results in excellent collection
efficiency. The detector is constructed of tissue equivalent plastic
and the filling gas is also tissue equivalent.
The instrument has been tested in fields of 10 Megarads per hour
instantaneous with excellent response. If you make beam on measurements
in nanosecond fields, you should consider the Model 1030.
Specifications
Mechanical
Dimensions:
4H, 5 1/4 W, 15 L
Weight: 5
1/2 lb.
Case: Painted
aluminum
External
Controls: Range Switch with positions OFF, Battery, 100,10,1,.1,0.01
rad/h full scale, Reset, 10,1, 0.1 mrad full scale
Detector
Type: Sealed
tissue equivalent ion chamber: Volume 150 CC, Multiple parallel
plates inside sealed aluminum can.
Wall material:
A 150 TE Plastic
Wall Thickness:
0.07 TE Plastic. Aluminum can: 0.065
Operational
Readout:
3 1/5 Taut band meter
Ranges: Ratemeter
Ranges: 0-100, 10, 1, 0.1, 0.01 rad/h. Integrate Ranges 0-10, 1,
0.01 mrad
Reproducibility:
±15%
Temperature Dependence: ± 15% 0 to 50 C
Humidity
Sensitivity: Some sensitivity to high humidity conditions.
Geotropism:
The instrument will momentarily read upscale or downscale when the
instrument is moved vertically or horizontally.
Warm up time:
1 minute
Batteries:
4 D cells 150 hours, 4 # 505 22.5V batteries for collection voltage
Internal
Controls: Adjustment for each range, zero, HIV battery test
Reset Switch
Option: Mounted below meter. |