model 6035 is an ion chamber detector. It is designed to be operated
with the 6000 series components for readout. The detector is a stand
alone component needing only 12 volts to operate. The electrometer
and high voltage power supply are included. The sensitive electrometer
is housed in a sealed housing to eliminate atmospheric and electrical
noise. The output is a series of pulses proportional to the radiation
field. This detector is designed for pulsed and steady state radiation
fields. Specifications
Mechanical
Size Overall:
9 in W x 9 in L x 12 in H excluding mounting bracket.
Construction:
Case is Aluminum. Electrometer housing is gasketed cast aluminum.
Desiccant is provided in housing for humidity control.
External
connectors: Multipin for power and signal.
Detector
Detector
type: Ion Chamber
Useful range:
6035: 1 mR/h to 1000 mR/h, 6035A: 10 mR/h to 10 R/h, 6035B: 0.1
mR/h to 100 mR/h
Sensitivity:
6035: 600 cpm/mR/h, 6035A: 60 CPM/mR/h, 6035B: 1200 CPM/mR/h
Suggested
Alarm setting range: 10 times lowest end of useful range
Energy response:
80 Kev to 1600 Kev within 30%
Environmental
Temperature
range: -10 to 50 C
Humidity
range: 0 to 99% non condensing